The Flash-Recorder-2 MITX self-recording voltmeter is a small-sized data collection system running under the Windows Embedded standard 7 operating system and is an industrial version of the Flash-Recorder-2-16/. It has the same technical characteristics, differs in the built-in processor to make the final turnkey solution according to buyer’s technical task.
The presence of a built-in processor allows you to create stands and automated systems based on it with information output to an external display connected linked to the HDMI connector.
Technical specification:
Characteristic |
Meaning |
DC voltage measure range (bipolar) |
From – 10 up to + 10 |
Limits of the permissible basic reduced accuracy of measuring DC voltage, %* |
± 0,1 (option ±0,05) |
Measuring range of the RMS value of the AC voltage, V |
From 0 up to 10 |
Limits of the permissible basic reduced error of measuring the RMS value of the AC voltage, %* |
± 0,1 (option ±0,05) |
Frequency of AC voltage, Hz |
From 0 up to 3000 |
DC power measuring range , мА |
From 0 up to 5 From 4 up to 20 |
Limits of the permissible basic reduced error of measuring DC power, %* |
± 1 (option ±0,5) |
Measuring range of time intervals, ms |
От 10 до 10000 |
Limits of the permissible basic relative error of measuring time intervals, % |
± 1 (option ±0,5) |
Input overload, V |
not more, than ± 40 |
Number of analog inputs (differential/multiplexed, with common wire) Flash-Recorder-2-16-MITX1 |
8/16 |
Number of digital input/output channels |
8/8 |
Number of synchronic ADC channels |
2 |
ADC resolution |
16 |
Maximum frequency of ADC conversion per channel, kHz |
5 |
Gain factors of the programmable amplifier |
1, 10, 100, 200 |
Input impedance, MOm, not less, than |
1 |
Bandwidth by minus level 3 dB, kHz, not less, than |
3 |
Insulation resistance between input and case, МОм, not less, than |
1 |
DC power supply voltage, V |
12 |
Communication interface with an external PC |
Ethernet |
Maximum distance from the computer to the devices, m |
500 |
Overall dimensions (length×width×height), mm |
160×150×50 |
Weight not more, than, kg |
3 |
Normal conditions of use: – temperature, °С – humidity, % – air pressure, kPa |
20 ± 5 up to 85 from 84,0 up to 106,7 |
Working conditions : – temperature, °С
– humidity, % – air pressure, kPa |
from + 5 up to + 50 (option from – 25 up to + 70) up to 90 from 84,0 up to 106,7 |
Average time to failure, h, at least |
10000 |
Average service life, years, at least |
7 |